Q-TRAK INDOOR AIR QUALITY MONITOR 7575 can be rented as part of a IAQ Test measurement. Q-Trak Indoor Air Quality Monitor Model 7575 provides quick, accurate information to assess key indoor air quality, or IAQ, parameters. Providing a comfortable, safe, and healthy indoor environment is an increasingly important concern. Good indoor air quality improves concentration and productivity. Good IAQ can also reduce lost days due to absence.
The Q-TRAK INDOOR AIR QUALITY MONITOR 7575 kit contain:
- Instrument
- Hard carrying case
- 4 alkaline batteries
- USB cable
- Trak-Pro software
- Operation and service manual
- AC adapter
- Calibration certificate
Applications
- IAQ investigations
- Industrial hygiene surveys
- Baseline trending and screening
- Building commissioning
- Tracking down emissions to their source (point source location)
Technical Specifications
Carbon Monoxide (IAQ Probe Model 982)
- Sensor Type Electro-chemical
- Range 0 to 500 ppm
- Accuracy5 ±3% of reading or 3 ppm, whichever is greater
- Resolution 0.1 ppm
- Response Time <60 seconds to 90% step change
Carbon Dioxide (IAQ Probe Models 980 and 982)
- Sensor Type Dual-wavelength NDIR (non-dispersive infrared)
- Range 0 to 5,000 ppm
- Accuracy6 ±3.0% of reading or ±50 ppm, whichever is greater
- Resolution 1 ppm
- Response Time 20 seconds
Temperature (IAQ Probe Models 980 and 982)
- Sensor Type Thermistor
- Range 0 to 60°C (32 to 140°F)
- Accuracy3 0.5°C (±1.0°F)
- Resolution 0.1°C (0.1°F)
- Response Time 30 seconds (90% of final value, air velocity at 400 ft/min [2 m/s])
Relative Humidity (IAQ Probe Models 980 and 982)
- Sensor Type Thin-film capacitive
- Range 5 to 95% RH
- Accuracy4 ±3% RH
- Resolution 0.1% RH
- Response Time 20 seconds (for 63% of final value)
% Outside Air
- Range 0 to 100%
- Resolution 0.1%
Barometric Pressure
- Range 517.15 to 930.87 mm Hg (20.36 to 36.648 in. Hg)
- Accuracy ±2% of reading
Operating Temperature
- 5 to 45°C (40 to 113°F)
Storage Temperature
- -20 to 60°C (-4 to 146°F)
Logging Capability
- Range Logs up to 56,035 data points with key (4) measured
- parameters enabled, 39 days at 1-minute log intervals
Time Constants
- 1 sec, 5 sec, 10 sec, 20 sec, 30 sec (user selectable)
Log Intervals
- 1 second up to 1 hour (user selectable)
Features
- Model 7575 simultaneously measure CO2, CO, temperature, and humidity
- Calculates dew point, wet bulb, and percent outside air
- Large graphic display
- Displays up to 5 measurements
- On-screen messages and instructions
- One instrument with multiple plug-in probe options including VOC’s and air velocity
- Store up to 39 days of data collected at one-minute log intervals
- TRAKPRO™ data analysis software provided for data logging, analysis and documenting results