The Pulsar Model 105 Class-1 Acoustic Calibrator may be rented as part of a complete noise measurement kit, containing all the requisite equipment to carry out field calibration for noise survey.
Pulsar Model 105 Class-1 Acoustic Calibratorkit contains:
- Pulsar Model 105 Class-1 Acoustic Calibrator
- Carrying case
- Batteries
- Calibration certificates
- User manual
Applications
- Acoustic Calibration of noise meters and dosimeters
Standards
- IEC 60942 Class 1
Technical Specifications
- Frequency : 1 kHz ± 1%
- Sound level : 94 dB re 20 μPa
- Standardisation model 105 : IEC 60942:2003 Class 1
- Distortion : Less than 2%
- Operating humidity : 25 to 90% Relative humidity
- Operating static pressure : 65 kPa to 108kPa
- Operating temperature : -10°C to +50°C
- Storing temperature : -20°C to +60°C
- Cavity diameter : 0.525 inch
- Battery : 1 x 9v 6F22 (Neda 1604)
- Battery life : Approx. 15 Hours continuous use
- Dimensions : 135 mm x Ø48 mm
- Weight with battery : 185g
Features
- Advanced performance ensures that both models meet or exceed IEC 60942 accuracy requirements
- Full PTB pattern approval
- Ergonomic and robust design ensures comfortable use
- Auto-correction for temperature and barometric pressure
- Auto switch off in normal mode and ‘permanent-on’ mode available
- 94dB Sound pressure level
- Fits ½ inch international standard microphones
- Calibration certificate included